Quality of tuberculosis (TB) microscopy diagnosis is not a guarantee despite implementation of external quality assurance (EQA) services in all laboratories of health facilities. Hence, we aimed at evaluating the technical quality and the findings of sputum smear microscopy for acid fast bacilli (AFB) at health centers in Hararge Zone, Oromia Region, Ethiopia. Of the total 55 health center laboratories assessed during the study period (July 2014-July 2015), 20 (36.4%) had major technical errors; 13 (23.6%) had 15 false negative results and 17 (30.9%) had 22 false positive results. The quality of AFB smear microscopy reading and smearing was low in most of the laboratories of the health centers. Therefore, it is essential to strength the EQA program through building the capacity of laboratory professionals.

The tuberculosis(TB) control program of Rwanda is currently phasing in light emitting diode-fluorescent microscopy (LED-FM) as an alternative to Ziehl-Neelsen (ZN) smear microscopy. This, alongside the newly introduced Xpert (Cepheid, Sunnyvale, CA, USA) is expected to improve diagnosis of TB and detection of rifampicin resistance in patients at health facilities. We assessed the accuracy of smear microscopy and the incremental sensitivity of Xpert at TB laboratories in Rwanda. This was a cross-sectional study involving four laboratories performing ZN and four laboratories performing LED-FM microscopy. A total of 96 presumptive pulmonary tuberculosis participants were culture positive for M. tuberculosis. The overall incremental sensitivity of Xpert over smear microscopy was 32.3 %; p 