The tuberculosis(TB) control program of Rwanda is currently phasing in light emitting diode-fluorescent microscopy (LED-FM) as an alternative to Ziehl-Neelsen (ZN) smear microscopy. This, alongside the newly introduced Xpert (Cepheid, Sunnyvale, CA, USA) is expected to improve diagnosis of TB and detection of rifampicin resistance in patients at health facilities. We assessed the accuracy of smear microscopy and the incremental sensitivity of Xpert at TB laboratories in Rwanda. This was a cross-sectional study involving four laboratories performing ZN and four laboratories performing LED-FM microscopy. A total of 96 presumptive pulmonary tuberculosis participants were culture positive for M. tuberculosis. The overall incremental sensitivity of Xpert over smear microscopy was 32.3 %; p 

Ziehl-Neelsen (ZN) bright-field microscopy is time-consuming, with poor sensitivity, even under optimal conditions. Introduction of Primo Star iLED fluorescent microscopy (FM) may improve TB case finding at referral hospitals in Rwanda. Laboratory technicians demonstrated high acceptance of iLED. The 100% concordance at both demonstration sites during the continuation phase shows technicians' rapid command of the new iLED microscope in a relatively short time. Additionally, fluorescent microscopy reduces the time necessary for examination by more than half. The high level of agreement between iLED and FM during implementation in both sites provides initial evidence for iLED to replace current methods.